Sensor and verified by the Reader. The communication between the Reader and Sensor for Alarm data
is a standard Bluetooth Low Energy (BLE) connection. The pairing of the Sensor to the Reader is
accomplished during activation with an authenticated login procedure that uses an Out-of-band key
exchange (NFC). This prevents unauthorized devices from connecting to the Sensor. The transmitted
data is protected by a proprietary data format and encryption. This prevents unauthorized devices from
accessing the data if they are within range and intercept the transmission. Under normal operation, the
industry standard BLE protocols allow for many users to be in the same vicinity. In the case where the
connection is lost due to out-of-range or interference, reconnection is only possible with the
authenticated Reader that activated the Sensor.
The communication between the App and Sensor during a scan is a short range Near Field
Communication (NFC) method which makes it di"cult to interfere with or intercept during
transmission. The transmitted data is protected by a proprietary data format, encryption, and memory
map. The integrity of transmitted data is ensured by a cyclic redundancy check (CRC) generated by the
Sensor and verified by the App. The communication between the App and Sensor for Alarm data is a
standard Bluetooth Low Energy (BLE) connection. The pairing of the Sensor to the App is accomplished
during activation with an authenticated login procedure that uses an Out-of-band key exchange (NFC).
This prevents unauthorized devices from connecting to the Sensor. The transmitted data is protected by
a proprietary data format and encryption. This prevents unauthorized devices from accessing the data if
they are within range and intercept the transmission. Under normal operation, the industry standard
BLE protocols allow for many users to be in the same vicinity. In the case where the connection is lost
due to out-of-range or interference, reconnection is only possible with the authenticated App logged in
to the same LibreView account that activated the Sensor.
Quality of Service (QoS):
QoS for the FreeStyle Libre 2 Reader and Sensor wireless communications using NFC is assured within
the e!ective range of 4 cm between the Sensor and Reader that is specified to occur within 15 seconds.
QoS for the wireless communication using BLE is assured between the Reader and Sensor at regular 1-
minute intervals. If connection is lost between the Reader and Sensor for 5-minutes, the connection lost
symbol displays. If connection is lost for 20 minutes, the Reader alarms the user if the alarm is turned
on. If connection is lost between the Sensor and the Reader, up to 8 hours of glucose results can be
retrieved by performing a scan with the Reader. The Reader is designed to only accept radio frequency
(RF) communications from recognized and paired Sensors.
QoS for the FreeStyle Libre 2 app and Sensor wireless communications using NFC is assured within the
e!ective range of 1 cm between the Sensor and phone. QoS for the FreeStyle Libre 2 app and Sensor
wireless communications using BLE is assured at regular 1-minute intervals. If connection is lost
between the App and Sensor for 5-minutes, the alarms unavailable symbol displays. If connection is lost
for 20 minutes, the App alarms the user if the alarm is turned on. If connection is lost between the
Sensor and the App, up to 8 hours of glucose results can be retrieved by performing a scan with the
App. The App is designed to only accept radio frequency (RF) communications from recognized and
paired Sensors.
Labeling Symbols