Descargar Imprimir esta página

Nouvag MD11 Instrucciones De Uso página 43

Ocultar thumbs Ver también para MD11:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 133
MD 11
6
Startup
6.1
Device setup
• Installation layout
• Place the MD 11 and all required accessories and instruments on an even, non-slip surface and make sure
you have good access to all controls.
• Do not allow the operating range of the device (including the cable and contra angle) to be compromised
by limiting factors.
• The system display must be fully visible at all times.
• The pedal must be placed within stepping distance between the patient and the surgeon.
• It must be explicitly ensured that no objects can fall on the pedal.
• The power plug at the rear of the device must be accessible at all times.
• The motor ventilation slots must be kept clear in order to prevent the motor temperature from
becoming excessive.
6.2
Connection to the power supply
Before plugging the power cable into the power socket for the first time, you must check the supply
voltage setting next to the power switch.
If the voltage shown does not correspond to the local mains voltage, the grey fuse holder must be set to the
correct voltage:
A)
Unplug the power cable.
B)
Use a screwdriver to open the fuse slot.
C)
Remove the fuse holder.
D) Remove the grey fuse holder and reinsert it so that the local mains voltage setting is shown in the
small window.
E)
Slide the fuse holder back in and close the fuse slot.
F)
Check the mains voltage shown on the fuse slot.
G) Plug the power cable back into the device.
In order to prevent the risk of electric shock, the device may only be connected to a power network with a
PE protective ground conductor.
MD 11, V2.0, Operating Instructions, REF 31665, V22/17
EN
7

Publicidad

loading