Descargar Imprimir esta página

Miracle-Ear CHARGE fullpower RIC/BTE T AV Manual Del Usuario página 9

Publicidad

Idiomas disponibles

Idiomas disponibles

Regulatory Information
We are required by regulations to provide the following warnings:
WARNING:
Use of the charger directly next to other electronic equipment
should be avoided because it could result in improper performance. If such use
is necessary, note as to whether your hearing aids and the other equipment are
operating normally.
WARNING:
Use of accessories, components or replacement parts other
than those provided by the manufacturer of the charger could result in increased
electromagnetic emissions and decreased electromagnetic immunity and could
result in degradation of performance.
WARNING:
If Portable Radio Frequency communications equipment
is used closer than 30 cm (12 inches) from your charger, degradation of the
performance of your charger could result. If this occurs, move away from the
communications equipment.
DO NOT OPEN CHARGER, NO USER-SERVICEABLE PARTS INSIDE
The service life of the charger is 3 years.
The Miracle-EarCHARGE
fullpower RIC/BTE T AV charger has a rating of IP52
per IEC 60529. This means that the Miracle-EarCHARGE
charger is protected from limited dust ingress and against water spray less than
15 degrees from vertical.
Do not connect the Miracle-EarCHARGE
devices not described in this manual.
16
fullpower RIC/BTE T AV
fullpower RIC/BTE T AV charger to any
Regulatory Information
The Miracle-EarCHARGE
fullpower RIC/BTE T AV charger has been tested to,
and has passed, the following emissions and immunity tests:
• IEC 60601-1-2 radiated and conducted emissions requirements for a
Group 1 Class B device as stated in CISPR 11.
• Harmonic distortion and voltage fluctuations affecting the power input
source as stated in Table 2 of IEC 60601-1-2.
• RF radiated immunity at a field level of 10 V/m between 80 MHz and
2.7 GHz as well as higher field levels from communications devices as stated
in Table 9 of IEC 60601-1-2.
• Immunity to power frequency magnetic fields at a field level of 30 A/m as
well as proximity magnetic fields as defined in table 11 of 60601-1-2.
• Immunity to ESD levels of +/- 8 kV conducted discharge and +/- 15 kV air
discharge.
• Immunity to Electrical fast transients on the power input at a level of +/- 2
kVat a 100 Hz repetition rate.
• Immunity to Surges on the power input of +/- 1 kV line to line.
• Immunity to conducted disturbances induced by RF fields on the power
input as stated in Table 6 of IEC 60601-1-2.
• Immunity to voltage dips and interruptions on the power input as stated in
Table 6 of IEC 60601-1-2.