Replace the Orbit
the soft cannula becomes kinked during insertion or if the can-
nula becomes fully or partially dislodged from the skin. Check
frequently to make sure the soft cannula remains firmly in place.
It must be completely inserted at all times to receive the full
amount of medication.
For insulin therapy, 2 – 3 hours after insertion, a blood glucose
test has to be performed to ensure adequate delivery of insulin.
Therefore, do not change the Orbit
ately before bedtime.
Develop a conventional insulin injection or medication replace-
ment plan with your healthcare professional if any part of the
infusion system fails or system supplies are depleted.
If these instructions are not followed, there is a risk of incorrect
delivery of medication. Seek immediate medical assistance if
you have any health concern.
Avoid bending or breaking the cannula/introducer needle.
A broken needle can remain stuck in the body or remain com-
pletely under the skin and may require surgical removal.
For Healthcare Professionals: Avoid recapping or destroying
the introducer needle. Recapping or destroying the intro-
ducer needle can lead to accidental sharps injuries which can
transmit serious infections. Healthcare professionals must fol-
low the specific directions of local regulations for healthcare
professionals using infusion sets, including directions for recap-
ping and disposing. Seek immediate medical assistance if you
receive a sharps injury.
Caution (please note)
• Orbit
soft Infusion Sets are designed to work with continuous
®
subcutaneous infusion pumps and pump reservoirs with stand-
ard luer lock connection. Follow the instructions for use of your
infusion pump and/or reservoir.
• Change the infusion site regularly in accordance with your
healthcare professional's recommendations. If this is not done,
you are at risk for infusion site infection, issues with medication
delivery, insulin resistance, and/or lipohypertrophy.
• Consult your healthcare professional to choose the correct can-
nula length. Your healthcare professional can provide recommen-
dations regarding the specific type of infusion set that is best
matched to your needs to ensure accurate delivery of medication.
• Do not put perfumes, deodorants or insect repellent on the
Orbit
soft Infusion Set as this may affect the integrity of the
®
infusion set. Protect the product from sunlight and heat. Store
Orbit
soft Infusion Sets in a dry location at room temperature
®
and avoid high humidity.
soft Infusion Set if the tape becomes loose, if
®
soft Infusion Set immedi-
®
11

Publicidad