Description; Product Overview - Dräger 200 ECO Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Description

CAUTION
!
Note the permissible temperature range from -15 to
+40 °C/+5 to+104 °F. The desired filling pressure may
not be achieved if the temperature is too high.
CAUTION
!
Risk of equipment damage in the non-permitted
pressure range of the gas supply.
If the gas supply is under too great a pressure or is
opened too quickly, the Dräger Oxygen Booster
200 ECO may be damaged.
Observe the permissible pressure range of 200 bar/
2900 psi (type A) or 207 bar/3000 psi (type B) for the
gas supply and always open all valves on the gas
supply and the Dräger Oxygen Booster 200 ECO
slowly.
NOTICE
i
i
Observe the documentation for the gas supply used
and the compressed air containers being filled.
NOTICE
i
i
The pressure ranges of the limit pressure switches
vary, as the type A and type B Dräger Oxygen Booster
200 ECO have different motor variants and therefore
have different limit pressure shutdown values.
2
Description
2.1

Product overview

A
1
2
3
4
5
Fig. 1
Overview
18
6
7
12
Fig. 2
Overview with optional filling panel
8
17
9
Fig. 3
Filling panel for six compressed gas cylinders (optional) in a
transport case
10
11
Item
Designation
1
G3/4 input connection (incl. sealing cap)
2
Inlet pressure gage
3
Rating plate
4
Filling valve
5
Filling pressure gauge
6
G3/4 output connection (incl. sealing cap)
7
Ventilation valve
8
Safety valve
9
On/off switch
10
Main switch
A
13
14
Dräger Oxygen Booster 200 ECO
15
16