7.10 Inspection and oiling
1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
– Damage to the coating.
– Bending of the product or parts of the product. Do not at-
tempt to straighten bent instruments!
2. Check the mechanical functionality of the product:
– Can the jaws be opened and closed?
– Can the handle be opened and closed?
– ERGO handle (20195-240): Can the upper part of the
handle be rotated freely relative to the lower part? Can
the locking knob be locked and loosened?
If damaged, do not use this product!
3. Oil all moveable parts of the handle and the joint area of the
insert using a physiologically acceptable oil (e.g. paraffin oil).
7.11 Packaging
The product can be sterilized while it is dismantled or assembled.
The jaws must be open.
The product must be protected against damage during steriliza-
tion.
1. If required, assemble the product. See page 24.
2. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/polyethylene and/
or in a sterilization container.
33