ERGO handle (20195-240): The upper part of the handle can be
rotated freely relative to the lower part. The required rotational
position can be set using the locking knob.
Do not coagulate unless the contact surfaces are within your
field of vision. Do not touch any metallic instruments during this
process.
6.7
Precleaning the product in the procedure room
1. Use a soft cloth (such as a compress) to remove any obvious
contamination immediately after use and/or place the prod-
uct in deionized water.
7
Cleaning, disinfection, sterilization
7.1
Safety instructions
The product must be cleaned and disinfected while it is disman-
tled and the jaws are open.
The product can be sterilized while it is dismantled or assembled.
The jaws must be open.
The instrument is also suitable for cleaning in an ultrasonic bath.
It is absolutely essential that the instrument parts lie on a non-
slip surface. The instrument parts must not come into contact
with other instruments, instrument parts, or the tank.
Clean, disinfect and sterilize the product again immediately after
use.
On no account use sharp or abrasive objects for cleaning purpos-
es.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
26