Blockage alarm check
Note: Alarm can be tested on either battery or AC
mains power. If testing on battery power, ensure battery
charge is sufficient: battery indicator is green.
1. Disconnect quick click adaptor from canister tubing
and cap off canister tubing quick click connector.
2. Reset stopwatch to zero.
3. Increase therapy set point to 100mmHg by pressing
Increase
icon.
4. Press Start/Pause Therapy
stopwatch.
5. Confirm that within approximately 3 minutes
touchscreen displays "WARNING BLOCKAGE",
status indicator illuminates yellow and audible alarm
sounds.
6. Press Start/Pause Therapy
device operation. Stop stopwatch.
7. If device fails test on first execution, check for any
damage to the canister and o-ring, ensure there
are no kinks in canister tubing and ensure canister
tubing quick click connector is capped. Re-test. If
device fails second time, replace o-ring and re-test.
If device fails again, remove from service and return
for repair.
Note: If device pump runs continuously at any time
during test (as determined by a quiet "putt-putt" sound),
then seal of test equipment is inadequate and test will
provide an inaccurate result, i.e., selected vacuum
levels not achieved within specified tolerances.
Restore presets
RENASYS™ TOUCH device settings must be restored to
factory presets between each patient use.
1. Press Setting
, scroll down and select Restore
Presets from Setting menu.
2. Press Accept
icon to restore device factory
presets.
3. Confirm:
• Alarm volume in Settings menu is set to HIGH.
• Compression rate in Settings menu is set to HIGH.
• Continuous therapy on Home screen is set to
80mmHg.
Note: Press the Therapy Mode toggle icon to switch
between Continuous
Annual service and maintenance
Replacement Components
Recommended personnel: Component replacement
button. Start
should be performed by a trained technician.
Equipment and materials required: Gloves, eye
protection, germicidal/antibacterial product, cotton
swab, RENASYS TOUCH O-ring, RENASYS TOUCH odor
filter, #1 Pozidriv® screwdriver / 5/32" (4mm) slotted
screwdriver, plastic o-ring removal tool.
Precautions:
• Use of gloves and eye protection or other personal
• Device should not be connected to AC mains power
button to stop
• Follow manufacturer's guidelines for safe use and
• Do not spray cleaning agent, immerse any part of
1. Remove existing o-ring using a plastic removal tool,
2. Ensure device is turned off and face down on a soft
3. Check that inlet port and air exhaust outlet are free
4. Apply germicidal/antibacterial product to a cotton
5. Dispose of cleaning materials in accordance with
6. Once inlet port is clean, carefully place new o-ring
7. Once air exhaust outlet is clean, carefully place new
8. Close access door. Tighten screw until door is flush
Caution: Do not over-tighten
10
• Intermittent therapy on Home screen set to
80mmHg for 5 minutes and 0mmHg for 2 minutes.
• Total Patient Therapy in Log is set to zero.
and Intermittent
and environmental protections is recommended.
while being cleaned.
disposal of cleaning materials.
device in fluid, or use an excessively wet cloth.
and dispose. Ensure inlet port is not damaged.
surface to avoid damage to touchscreen. Loosen
screw in rear access door. Remove and dispose of
existing odor filter.
of obstructions, dust or foreign materials.
swab and clean both inlet port and air exhaust
outlet.
facility protocol.
onto inlet port. Ensure o-ring is properly seated in
the groove.
odor filter behind access door.
with case.
Therapy

Publicidad

loading

Este manual también es adecuado para:

668012806680128166802134