Steps
Functionality check
20
External power indicator illuminates green after
power is connected.
21
The status indicator illuminates green when Power
button is pressed, Smith & Nephew startup screen
the device issues an audible tone and the welcome
screen displays.
22
Touchscreen is ready to accept user inputs.
23
Lightning bolt present in battery indicator to indicate
charging.
24
Record Serial Number, Operating System,
Main Application Version, Motor Controller Version,
Bootloader Version, RENASYS™ TOUCH System Version,
Battery Remaining, Device Lifetime Therapy and
Maintenance Due.
25
Device is set to Clinician mode.
Operation check
26
Status indicator illuminates green, therapy indicator
rotates orange, device performs leak check and
Delivering Therapy screen displays.
27
User interface locked.
28
User interface unlocked.
Operation on Battery check
29
Device operates on battery power.
Blockage alarm checks
30
Touchscreen displayed "WARNING BLOCKAGE",
status indicator illuminated yellow and audible alarm
sounded within approximately 3 minutes.
Yes No
Comments
Serial Number
Operating System
Main Application
Motor Controller
Bootloader System Version
RENASYS TOUCH System Version
Battery Remaining
Device Lifetime Therapy
Maintenance Due
14