RENASYS™ TOUCH service and maintenance sample checklist
Note: In the event that a device fails any test, has sustained obvious damage, has a missing or illegible label, or has
any indication it may not be fit for intended use, please return the unit to Smith & Nephew Service and Repair Center.
Service and maintenance sample check list UNIT SERIAL #
Steps
Device cleaning
1
All external surfaces of device, clamp, power cord
and power supply cleaned.
Device physical appearance check
2
No evidence of device damage, cracks or missing
pieces.
3
No evidence of loose parts inside the device.
4
Device touchscreen is not cracked, shattered or in
dismounted condition.
5
Device push buttons (Power, Start/Pause Therapy
and Lock/Unlock) are intact, physically undamaged
and symbols legible.
6
Device attachment knobs are intact.
7
Status indicator is not cracked, shattered, or in
dismounted condition.
8
External power indicator is not cracked
shattered or in dismounted condition.
9
Device power jack is physically undamaged.
10
Device identification label is present, REF and SN
numbers clearly visible.
11
No evidence of inlet port damage. O-ring is present.
12
Barcode label is intact and legible.
13
Power supply DC cord is not fraying, missing
insulation, and has no bent, loose or missing plug
blades or earth/ground blade.
14
Front and rear labels of power supply are legible.
15
Power cord is not fraying, missing insulation, and has
no bent, loose or missing plug blades.
16
No evidence of clamp damage, cracks or missing
pieces.
17
Clamp label is intact and legible.
18
Clamp clips open and close.
19
Clamp knob fully opens and closes arm.
Yes No
Comments
13