Steps
Restore presets
31
Alarm volume:
High
Compression rate:
High
Continuous therapy: 80mmHg
Intermittent therapy: 80mmHg for 5 min
0mmHg for 2 min
Total Patient Therapy: 0
Annual service and maintenance
32
O-ring replaced with new o-ring.
33
Odor filter replaced with new odor filter.
34
The vacuum measured by the pressure gauge is
25 ± 5mmHg when device set @ 25mmHg.
35
The vacuum measured by the pressure gauge is
200 ± 5mmHg when device set @ 200mmHg.
36
Entered maintenance mode and reset annual timer
Storage and battery maintenance
37
a B
e t t
y r
c
a h
g r
l e
v e
l e
s i
e b
Battery fully charged.
Cleaning and Packing
38
Device and power supply cleaned and packaged
with all necessary items for customer use or return.