Performing CPR (Cardiopulmonary Resuscitation)

1. Disconnect the air supply connector by grasping the grip ring and pulling it away from the
air supply unit.
Grip ring
i
to commence.
2. Swing the air supply hose away from the therapy unit.
3. Level the bed.
4. Begin CPR.
When the air supply connector is disconnected a CPR alarm screen will be displayed and the unit
will go into standby mode.
After CPR has been performed and the patient is stable
1. Raise the side rails and lock into position.
2. Reconnect the air supply connector to the therapy unit.
3. Reset the alarm condition to resume therapy.

Control Screen Display

will damage the screen and affect future use.
Pull ring away
Connector disconnects