Performing Cpr (Cardiopulmonary Resuscitation); Control Screen Display - arjo TheraKair Visio Manual Del Usuario

Ocultar thumbs Ver también para TheraKair Visio:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 222

Performing CPR (Cardiopulmonary Resuscitation)

1. Disconnect the air supply connector by grasping the grip ring and pulling it away from the air
supply unit.
Grip ring
i
It will take approximately 10 seconds for the mattress to deflate sufficiently for CPR
to commence.
2. Swing the air supply hose away from the therapy unit.
3. Level the bed.
4. Begin CPR.
When the air supply connector is disconnected a CPR alarm screen will be displayed and the unit will go
into standby mode.
After CPR has been performed and the patient is stable
1. Raise the side rails and lock into position.
2. Reconnect the air supply connector to the therapy unit.
3. Reset the alarm condition to resume therapy.

Control Screen Display

The control screen display on the front of the unit is a touch sensitive screen. The system control screens
will be displayed on this screen. These screens will display information on current system operations and
settings and allow the alteration of these values.
The screens have soft buttons; this means that the buttons operate exactly as real buttons would, simply
press the buttons to activate them.
The operation of the control screens is detailed in the following pages.
The control screen should only be operated by finger; using pens or pointing devices will
damage the screen and affect future use.
Pull ring away
19
Connector disconnects

Publicidad

Tabla de contenido
loading

Tabla de contenido