4 Mounting
Mounting socket
14
EN 61326: class A). In this case, use a probe with coaxial version.
Fig. 5: Mounting in non-metallic vessel
1
Flange
2
Metal sheet
If this is not possible, use short sockets with small diameter.
Higher sockets or sockets with a bigger diameter can generally be
used. They can, however, increase the upper blocking distance (dead
band). Check if this is relevant for your measurement.
In such cases, always carry out a false signal suppression after
Fig. 6: Mounting socket
vessel top.
1
d
DN25 ... DN150
> DN150 ... DN200
d
2
Setup procedure".
h
≤ 150 mm ( 5.91")