Caution:
When mounting, make sure that the cable is straight over its entire
length. A kink in the cable can cause measurement errors and contact
with the tube.
With rod probes, a spacer is generally not required. However, if there
is a risk of the rod probe being pressed against the tube wall by in-
contact with the tube wall. In the case of cable probes, the cable can
be strained.
Keep in mind that buildup can form on the spacers. Strong buildup
Note:
Measurement in a standpipe is not recommended for extremely
adhesive products. In case of slight buildup, you should choose a
standpipe with bigger diameter.
Instructions for the measurement:
The 100 % point with standpipes should be below the upper
ventilation hole.
The 0 % point in standpipes should be above the gravity or center-
ing weight.
A false signal suppression with installed sensor is generally rec-
ommended to achieve the best possible accuracy.
If there is a risk of the cable probe touching the vessel wall during
operation due to product movements or agitators, etc., the measuring
There is an internal thread (M8) in the gravity weight, e.g. for an eye-
bolt (optional).
Make sure that the probe cable is not completely taut. Avoid tensile
loads on the cable.
this condition can lead to measurement errors.
If there is a danger of the rod probe touching the vessel wall, fasten
the probe at the bottom end.
Keep in mind that measurement is not possible below the fastening
point.
4 Mounting
19