Diagnostics - Echo curve
Diagnosis - Simulation
In another window you can carry out a reset of the two peak values
separately.
Information:
available.
The menu item "Echo curve" shows the signal strength of the echoes
over the measuring range in V. The signal strength enables an evalua-
tion of the quality of the measurement.
curve.
"X-Zoom": Zoom function for the meas. distance
In this menu item you can simulate measured values via the current
output. This allows the signal path to be tested, e.g. through down-
stream indicating instruments or the input card of the control system.
Select the requested simulation variable and set the requested value.
Caution:
During simulation, the simulated value is outputted as 4 ... 20 mA cur-
rent value and digital HART signal.
6 Set up with the display and adjustment module
V"