Telstar Bio II Advance Plus Serie Manual De Usuario página 36

Cabinas de seguridad microbiológica
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

No.
Type of
Causes of error
error
Excessive
Probe not working properly.
laminar
down flow
3
Alarm "
HIGH
Air currents in front
LAM. FLOW
"
aperture.
(see Chapter
4.5 point D)
Probe not working properly.
Laminar
down flow
not working
properly
4
Alarm "
LOW
LAM. FLOW
"
(see Chapter
Laminar down flow fan
4.5 point D)
stopped.
Exhaust completely covered
up.
Exhaust not
Probe not working properly
working
properly
5
Alarm "
LOW
INFLOW"
(see Chapter
4.5 point D)
Exhaust fan stopped.
Front does
6
Front is blocked.
not move
Glass position
Glass out of
position
Working position micro is
Alarm
loose
"
WORK
7
POSITION"
(see Chapter
4.5 pto. D)
Main control is damaged
UM-BioIIAdvancePlus-EN-1812
Correction of error
Check the probe position is correct (sensor
parallel to the flow).
Check and remove any airflow in front of
the cabinet that distorts the reading of the
sensor.
Check the probe position is correct (sensor
parallel to the flow).
Check and remove any obstruction to the
probe.
Check that the connection in the control
card is correct (Annex A7).
Check the fuse has not blown (Annex
A7).
Check that the connection in the control
card is correct (Annex A7).
Notify the Technical Service to change the
fan.
Visually check the exhaust to remove any
type of obstruction.
Check the probe position is correct (sensor
parallel to the flow).
Check that the connection in the control
card is correct (Annex A7).
Check the fuse has not blown (Annex
A7).
Check that the connection in the control
card is correct (Annex A7).
Notify the Technical Service to change the
fan.
Visually check the front movement to
remove any type of obstruction.
Assemble the glass in working position.
Open the frontal and check that the
working position micros are not loose.
Notify the Technical Service to receive
instructions for adjusting the micros.
Call a specialised technician to replace the
control.
34 / 35

Publicidad

Tabla de contenido
loading

Tabla de contenido