TECHNICAL FEATURES

Features
External dimensions (LxWxH)
Internal dimensions (LxWxH)
Front opening height
Weight
Downflow velocity
Inlet velocity
Consumption*
Voltage
Frequency
Lighting
Noise
Vibration
HEPA/ULPA H 14 Filters
Fans
* Cabinet with factory settings and new filter.
DISSOCIATED DECLARED VALUES OF NOISE EMISSION
Standard(s) to which conformity is declared ISO 4871
Bio II Advance Plus 3
L
(dB)
pA
K
(dB)
pA
Values measured according to acoustic test in standard EN 12469:2000 using standard
UM-BioIIAdvancePlus-EN-1812
Units.
Bio II
Advance Plus
1049 x 764 x
mm
954 x 602 x
mm
mm
Kg
m/s
m/s
W
V
Hz
Lux
dB (A)
mm RMS
Efficiency 99.995%, EN 1822 by MPPS method
filtration system in laminar and exhaust flow
(99.999% D.O.P. test @ 0.3µm particles)
V
= 0.35m/s
downflow
V
= 0.55 m/s
inlet
(230 V, 50 Hz)
Bio II Advance Plus 4
≤58
2
ISO 11201
Bio II
Advance Plus
3
4
1354 x 764 x
1260
1260
1259 x 602 x
580
580
182
180
200
0.35
0.35
0.55
0.55
125
125
230
50/60
≥ 1000
≤ 58 (Attached table)
< 0.005
Two high-efficiency fans (four in cabinet
Advance Plus 6). Flows measured by a
thermoanemometric probe.
Bio II Advance Plus
≤58
2
Bio II
Advance Plus
6
1964 x 764 x
1260
1869 x 602 x
580
280
0.35
0.50
338
V
= 0.35m/s
downflow
V
= 0.50 m/s
inlet
(230 V, 50 Hz)
6
≤58
2
7 / 35