Technical Features - Telstar Bio II Advance Plus Serie Manual De Usuario

Cabinas de seguridad microbiológica
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

2.6.

TECHNICAL FEATURES

Features
External dimensions (LxWxH)
Internal dimensions (LxWxH)
Front opening height
Weight
Downflow velocity
Inlet velocity
Consumption*
Voltage
Frequency
Lighting
Noise
Vibration
HEPA/ULPA H 14 Filters
Fans
* Cabinet with factory settings and new filter.
DISSOCIATED DECLARED VALUES OF NOISE EMISSION
Standard(s) to which conformity is declared ISO 4871
Bio II Advance Plus 3
L
(dB)
pA
K
(dB)
pA
Values measured according to acoustic test in standard EN 12469:2000 using standard
UM-BioIIAdvancePlus-EN-1812
Units.
Bio II
Advance Plus
1049 x 764 x
mm
954 x 602 x
mm
mm
Kg
m/s
m/s
W
V
Hz
Lux
dB (A)
mm RMS
Efficiency 99.995%, EN 1822 by MPPS method
filtration system in laminar and exhaust flow
(99.999% D.O.P. test @ 0.3µm particles)
V
= 0.35m/s
downflow
V
= 0.55 m/s
inlet
(230 V, 50 Hz)
Bio II Advance Plus 4
≤58
2
ISO 11201
Bio II
Advance Plus
3
4
1354 x 764 x
1260
1260
1259 x 602 x
580
580
182
180
200
0.35
0.35
0.55
0.55
125
125
230
50/60
≥ 1000
≤ 58 (Attached table)
< 0.005
Two high-efficiency fans (four in cabinet
Advance Plus 6). Flows measured by a
thermoanemometric probe.
Bio II Advance Plus
≤58
2
Bio II
Advance Plus
6
1964 x 764 x
1260
1869 x 602 x
580
280
0.35
0.50
338
V
= 0.35m/s
downflow
V
= 0.50 m/s
inlet
(230 V, 50 Hz)
6
≤58
2
7 / 35

Publicidad

Tabla de contenido
loading

Tabla de contenido