5.1.2

Cleaning

The outer case and display may be wiped with a clean towel lightly moistened with 70% isopropyl alcohol (IPA) or a 10% bleach solution.
Do not introduce the cleaning solution or any other liquids directly into the unit. Do not use a saturated towel which may introduce liquid into
the case or display seams. Ensure that the BD Veritor Plus Analyzer is dry and the surface is free of any residual cleaning solution prior to
returning to use.
If the BD Veritor InfoSync or BD Veritor InfoScan module is installed, BD does not recommend cleaning the barcode scanner window with
any cleaning agent. Use a clean, soft towel lightly moistened with plain water to clean the window gently. Scratching the window may
reduce the scanner's performance.
5.2

Servicing

There are no user serviceable components in the unit. For technical issues or questions, please contact your local BD representative.
5.3

Transportation and Storage

The Analyzer and optional modules should be transported in the original packaging
14 days, up to 85% RH, non-condensing. BD recommends the original packing materials be retained for this purpose. Follow local
transportation regulations for shipping equipment containing a Li-ion secondary battery.
The BD Veritor Plus Analyzer should be stored at 15–30 °C, up to 85% RH, non-condensing.
6.1

Instrument Service

If your BD Veritor Plus Analyzer malfunctions or operates unusually in any way, you may initially attempt to solve the problem by following
the recommendations in this section. All other servicing attempts will terminate the responsibility of the manufacturer under the terms of the
warranty.
If you cannot repair a system malfunction, contact your local BD representative (contact numbers are listed on page 191).
6.2

Forced Power Down

If the BD Veritor Plus Analyzer is not responsive and does not turn off using the standard power down procedure, hold the power button
down for approximately 10 seconds. The Analyzer will power down. Normal operation may then be resumed by following the standard
power on sequence.
6.3

Troubleshooting Aids

The barcode scanner in the BD Veritor InfoScan Module or BD Veritor InfoSync Module may be used to initiate the display of system
information or to enter a barcode scanner test mode.
Analyzer Information Display Mode
This mode will display the following system information:
Analyzer serial number
Analyzer expiration date and remaining test count
Current configuration settings
Firmware version
6 – Troubleshooting
S/N: 123456789
Expiration: DD-MMM-YYYY
Tests Remain:
OID: No
SID: Yes
LOT: No
Language: English
FW Ver:1.00
-20–65 °C for up to 2 days, -20–45 °C for up to