Result Number
Message
SPECIMEN ID
SCAN TIMEOUT
none
REINSERT
TEST DEVICE
LOT NUMBER
SCAN TIMEOUT
none
REINSERT
TEST DEVICE
XXX DAYS DATA
none
SUBSCRIPTION
REMAINING
XXX DAYS
none
ANALYZER LIFE
REMAINING
XXX READS
none
REMAINING
none
SUBSCRIPTION
EXPIRED
Flashing Plug icon
none
WALK AWAY MODE
REMOVE TEST DEVICE
none
INSERT
WHEN PROMPTED
TEST DEVICE
none
WAS REMOVED
REPEAT ASSAY
Possible Cause(s)
No barcode was scanned
within 30 seconds.
No barcode was scanned
within 30 seconds.
The service agreement to
transmit results from the
BD Veritor Plus System to
the local informatics system
is nearing expiration.
The shelf or operational life
of the BD Veritor Plus
Analyzer is nearing
expiration. The maximum
shelf-life of the BD Veritor
Plus Analyzer from the date
of manufacture is 34 months
(regardless of time of first
use). The maximum number
of months from first use is
24 months.
The BD Veritor Plus Analyzer
is nearing its usage
expiration. The maximum
number of tests performed
by a BD Veritor Plus
Analyzer is 3500 runs.
The service agreement to
DATA
transmit results from the
BD Veritor Plus System to
the local informatics system
has expired.
Failure of the internal battery.
Test device inserted at the
incorrect workflow step
during Walk Away operation
mode.
Test device was removed
before the test run was
complete during Walk Away
operation mode.
Corrective Action(s)
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
Contact your BD sales representative or BD Customer
Service.
The duration of the message is two seconds. To clear the
message immediately, insert a test device to begin a new
run.
Contact your distributor to order a new BD Veritor Plus
Analyzer.
The duration of the message is two seconds. To clear the
message immediately, insert a test device to begin a new
run.
Contact your distributor to order a new BD Veritor Plus
Analyzer.
The duration of the message is two seconds. To clear the
message immediately, insert a test device to begin a new
run.
Contact your BD sales representative or BD Customer
Service.
Note: the Analyzer may continue to be used and results
read from the display window. Test results will be queued
in the Analyzer memory and transmitted once the service
agreement is renewed.
Contact BD.
Note: the Analyzer may continue to be used while
plugged into a facility power source.
Current run aborted. Remove test device and
immediately restart the Walk Away test sequence. Insert
the test device after the display window prompt to do so.
Current run aborted. The test must be repeated with a
new test device because the elapsed development time
of the assay is unknown.