BD Veritor™ Plus Analyzer Instructions for Use
Cellular Technology
(BD Veritor InfoSync Module only)
Data Transmission Security Protocols
Default Workflow Configuration
Operator Identification Scan
Specimen Identification Scan
Kit Lot Information Scan
Display Language
USB Port
2.6

Module Setup

Optional
The BD Veritor InfoScan package includes a USB cable which may be used to download test results to a computer.
A BD Veritor Plus Analyzer can accommodate either one BD Veritor InfoScan Module or one BD Veritor InfoSync Module. The steps for
installing the module are the same for both:
1
Turn the Analyzer power off.
2
Turn the Analyzer upside down.
Ensure that the power button is not accidentally pressed while the Analyzer is upside-down.
3
Depress the locking tab on the module bay cover.
ELECTRICAL EXPOSURE HAZARD. DO NOT TOUCH THE ELECTRICAL CONTACTS INSIDE THE MODULE BAY.
4
Slide the module bay cover off of the Analyzer. The module bay cover should be stored in a secure location.
5
Align the module alignment features to the Analyzer track.
6
Slide the module into the Analyzer until an audible click is heard.
7
The Analyzer is now ready for operation. See 4.6 Workflow Configuration Procedure – Optional for optional configuration instructions.
Steps 1–7 above complete the physical installation of the BD Veritor InfoSync Module,
but as the InfoSync Module is used to transmit assay results to a BD-hosted cloud server
and then to a facility's information system, additional IT setup is required. To create or troubleshoot
8
CAUTION
Power to the Analyzer should remain off during module installation.
WARNING
this connection, please contact your local BD representative.
Five band 3G HSPA
HTTPS with SSL/TLS certification. AES-256 payload encryption.
Disabled
Enabled
Disabled
English
Locked
NOTE