5.2.6. Dielectric Discharge
This measurement method is generally used to diagnose multilayer
insulation, which requires the instrument to measure the discharge
current and the capacitance of the object measured 1 min after the test
voltage is removed. This is a very good insulation diagnostic test to detect
deterioration and other problems in the multiple insulation to be
evaluated.
Table of reference values for analysis of the DD test result for
multilayer insulation
DD Result
• Access the MENU using the
• Use the control
• Use the control
Parameters
• Use the control
Dielectic Discharge.
Quality of insulation
> 7
4 - 7
2 - 4
< 2
button.
, select SETUP TEST and press
MODE SELECTION
SETUP TEST
LANGUAGE
, select D.D. TEST and press
=
SETUP TEST========
TEMPORIZED TEST
STEP VOLTAGE TEST
RAMP TEST
D.D. TEST
CAPACITANCE TEST
, select ON to activate or OFF to deactivate the