Descargar Imprimir esta página

MEGABRAS MD5KVR Manual De Uso página 46

Publicidad

Idiomas disponibles

Idiomas disponibles

12. Technical Specifications
Test voltages
Terminal output max value
Maximum resistance reading
DC Voltmeter
AC voltmeter
Over voltage protection
Current measurement
Capacitance measurement
Short circuit current
Test voltages accuracy
:
500 V, 1,000 V, 2,500 V, 5,000 V with fast
selection. Other test voltages from 50 V to 5
kV, can be selected in steps of 25 V or
500 V.
DC, negative in relation to grounding.
:
5,100 V.
:
5 TΩ
@ 5 kV.
1 TΩ
@525 V up to 999 V.
500 GΩ @500 V.
250 GΩ @ 250 V.
:
15 V up to 1,000 Vdc.
Precision: ± (5% of the reading + 3 digits)
:
15 V up to 1,000 V r.m.s.
Precision: ± (5% of the reading + 3 digits)
:
CAT. III – 600 V.
:
1 nA up to 3,000 μA.
± (10% of the reading + 3 digits)
:
50 nF up to 10 µF
50 nF up to 5 µF
30 nF up to 2 µF
30 nF up to 1 µF
Precision: ± 10% of the reading ± 3 digits
:
Max. 3 mA.
:
± 3% of nominal value over on 10 GΩ
resistance.
47
@ 500 V.
@ 1,000 V.
@ 2,500 V.
@ 5,000 V.