Example 2) Set up values for x:y Scaling = 0 to 100%
Settings
Settings
1
x:y
x:y
mA
%
+04.000
+04.000
=
+04.000
=
+20.000
+20.000
=
= +20.000
Sk1
Sk1 = Edit
= Flow scaling (mA, pressure only)
2. Operation
This section gives examples of how to connect and use the
instrument. Before you start:
•
Read and understand section "Safety" on page 1.
•
Do not use a damaged instrument.
2.1 Pressure Connections
CAUTION
To prevent damage, do not apply
torque to the body of the instrument. If
available, use the flat faces on the pressure
connector to hold the instrument in position.
Use an applicable method to seal the pressure
connections, and then tighten to the applicable torque
(Figure 1).
Note: If the instrument uses G1/8 connections, and the
pressure is ≥ 1500 psi (100 bar), connection method (b) is
mandatory.
a
2
1
3
c
1
2
4
a
1/8 NPT connections.
b
G1/8 connections: Pressure ≥ 1500 psi (100 bar)
c
G1/8 connections: Pressure < 1500 psi (100 bar)
d
M5 or 10-32 UNF connections: Refer to Table 9.
1
Process connector. Maximum torque:
1/8 NPT: 26 lbf.ft (35 Nm)
G1/8: 18.4 lbf.ft (25 Nm)
2
Applicable instrument pressure connector (Table 9).
3
(1/8 NPT only) Thread with an applicable sealant.
4
(G1/8 only) Applicable bonded seal.
5
Reference port connector: M5 or 10-32 UNF (Table 11)
Maximum torque: 1.5 lbf.ft (2 Nm)
Figure 1: Connection Methods
2.2 Communications Port Connections
Use the communications port (Figure A1 - item [9]) to
attach an IDOS Universal Measurement Module (UMM).
When you attach the cable from a UMM (Figure 8), the
instrument automatically changes the menus to give you all
the applicable options (Table 2 & Table 3).
2
Edit
3
4
= +00.000
MENU
OK
+
= +100.00
b
1
2
4
d
2
5
2.3 Measure Pressure
To measure the pressure or the pressure difference (if
there are two IDOS pressure ranges).
1. Connect the instrument (Figure 2) and, if necessary,
adjust the Set Up (Table 1).
2. Select an applicable pressure task from Select Task
(Table 2 & Table 3) and, if necessary, adjust the
Settings (Table 4).
3. If necessary, do a zero correction (Table 4).
4. Before you disconnect the instrument, isolate and
bleed the system.
p1
p2
P1
P2
P1 30 ... g
P1 ... g
P
30 ... g
MENU
ESC
OK
HOLD
20 ... g
a) P1 and/or P2
Figure 2: Example Configuration - Measure Pressure
The examples in Figure 2 shows:
•
How to measure one or two pressures in a system.
•
How to measure pressure difference.
Note: ∆P = High range IDOS - Low range IDOS. If
Pressure (p2) > Pressure (p1), the result is negative.
2.4 Leak Test
To do a leak test on a pressure system:
1. Connect the instrument (Figure 3) and, if necessary,
adjust the Set Up (Table 1).
2. Select an applicable leak test from Select Task
(Table 2 & Table 3) and, if necessary, adjust the
Settings (Table 4).
3. Set the period for the leak test (Table 4).
4. If necessary, do a zero correction (Table 4).
5. To start the leak test, select Start (). When the test
is finished, the instrument calculates the leak rate in the
applicable units/minute.
English–DPI800/802 Instruction Manual | 5
p1
p2
P
MENU
MENU
OK
OK
P1-P2
15.006
5.003
10.003
Settings
Settings
b) P1 - P2
Copyright 2003 Baker Hughes Company.