Routine check
XVII
1.
Functional test
a. Motor running?
b. Is vacuum sufficient?
ok?
2.
Open device
3.
Check QuatroFlex
– Film hinge torn
– Open valve cover
ok?
4. Check sound
– Dirt, discoloration
– Brittle material
ok?
5. Check tubings
– Check seal of all connections
– Dirt, discoloration
– Brittle material
ok?
6.
Close device
7.
Functional test
10
EN
(time required approx. 30 minutes)
not OK
a. Disconnect the pump from the main power supply
b. Release four screws
c. Remove rear panel
not OK
absorber (See diagram on page 9)
not OK
not OK
a. Clean interior of device
b. Attach rear panel
c. Tighten four screws
d. Plug in the main power cord
e. Testing of rechargeable battery (See Battery operation – page 8)
(See Functional test – page 7)
Motor not running
Contact your service center
Vacuum insufficient
Contact your service center
Replace QuatroFlex
Medela part # 077.1700
Replace sound absorber
Medela part # 077.1702
Replace QuatroFlex
Medela part # 077.1700