Descargar Imprimir esta página

Medela Invia Vario 18 AC/DC c/i Instrucciones De Uso página 8

Ocultar thumbs Ver también para Invia Vario 18 AC/DC c/i:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 163
Important notes
IX
In the application area of medical suction devices, the vacuum is normally quoted as the diffe-
rence (in absolute figures) between the absolute pressure and the atmospheric pressure or as
negative values in kilopascals (kPa). The indication, e.g. of -10 kPa, in this document therefore
always refers to a pressure level in kPa below the ambient atmospheric pressure.
Rechargeable battery: charge the battery before use. (See Battery operation – page 8)
Required maintenance
X
• Use equipment only under supervision of qualified medical personnel.
• Service Vario after each patient for safe and effective use. Service procedures can
be found on page 6.
• QuatroFlex Lifetime Rating: 4000 hours
• A device to determine pump usage hours is available from Medela.
• In case of excessive noise, consider replacing QuatroFlex.
• Vario Lifetime Rating: 2 years
• Contact your equipment supplier with any questions.
Service procedures
XI
6
EN
1. Check general condition of the device
• Is the overflow protection / bacteria filter present and
inserted correctly?
(See Filter test – page 7)
Change the filter... • once a week
• Check seal of tubing connections
• Suction jars and lids – Check for cracks, brittle areas,
discolouration? – exchange, if necessary
• Mechanical overflow protection underneath canister lid must
be open. To open pull downward.
• after an overflow
• when there is a change of patient

Publicidad

loading