Technical Documentation (Con't)
XXIX
IEC 60601-1-2, Table 206
Recommended separation distance between portable
and mobile RF communications equipment and the Medela
Suction Pumps
The Medela Vario 18 AC/DC c/i are intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the Medela Vario
18 AC/DC c/i can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Medela
Vario 18 AC/DC c/i as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation di-
stance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by ab-
sorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √ P
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d = 1.2 √ P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3 √ P
0.23
0.73
2.3
7.3
23