Disinfection Procedure:
• Always wear suitable work clothing, protective gloves and goggles
• Unpack equipment and place it on a suitable bench (sink)
• Open device and dispose of any soiled, contaminated parts like internal tubing,
valve plates, etc. according to local regulations
• Disassemble piston/cylinder units which are not worn out.
All possibly contaminated parts:
1. Spray the surfaces with hospital spray disinfectant and leave for 3 minutes
2. Clean the surfaces with a good cleaning agent, methodically rinse with water
3. Again spray the exterior with disinfectant and wipe parts dry with household
tissue or allow to air dry
• Dispose of protective gloves after the disinfection procedure
• Wash your hands
Service work
XXIII
•
Disconnect the pump from the main power supply
•
Remove four screws
•
Remove cover
A
Replacing the
rechargeable
battery
part # 077.0833
B
Replacing
QuatroFlex
part # 077.1700
C
Set interval
times
•
Attach cover
•
Insert four screws
•
Plug the pump in
•
Charge the rechargeable battery for five hours
• Functional Test (See Functional test – page 7)
All other service work must be performed by Medela or by an authorized Medela
service center.
A1 Detach the battery cable
A2 Remove the rechargeable battery
A3 Insert new rechargeable battery
A4 Attach battery cable
B1 Remove the three tubings
B2 Turn QuatroFlex 45° to the left
B3 Remove QuatroFlex
B4 Clean inside of the pump
B5 Insert QuatroFlex
B6 Turn QuatroFlex 45° to the right
B7 Re-attach all the tubings
C1 Determine the slot required
C2 Attach jumper in required
position
On
Off
8 Secs.
4 Secs.
16 Secs.
8 Secs.*
32 Secs.
16 Secs.
*factory setting