EMC DECLARATION
(In accordance with standard en 60601-1-2, november 2001 edition)
Table 201 - Guidance and Manufacturer's Declaration - Electromagnetic Emissions
G8 Evolution surgical lights are intended for use in the electromagnetic environment specified below. G8 Evolution customers or
users should ensure that they are used in such an environment.
Immunity test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage
fluctuations
and
emissions
IEC 61000-3-3
Table 202 - Guidance and manufacturer's declaration - electromagnetic immunity
G8 Evolution surgical lights are intended for use in the electromagnetic environment specified below. G8 Evolution customers or
users should ensure that they are used in such an environment..
IEC 60601
Immunity test
Electrostatic
± 6 kV contact
discharge
± 8 kV air
(ESD) IEC 61000-4-2
Electrical fast
± 2 kV for power supply
transient / burst
lines
IEC 61000-4-4
± 1 kV for input/output
lines
Surge
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
Voltage dips, short
< 5% U
interruptions and
(dip > 95% of U
voltage variations on
for 0.5 cycles
power supply lines
40% U
IEC 61000-4-11
(dip = 60% of U
for 5 cycles
70% U
(dip = 30% of U
for 25 cycles
< 5% U
(dip = 95% of U
for 5 cycles
Power frequency
3 A
(50/60 Hertz)
magnetic fields
IEC 61000-4-8
Note: U
is the voltage of the AC supply network before applying the test level.
T
30
Compliance
Group 1
Class A
Not applicable
flicker
Not applicable
Compliance
test level
level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV differential
mode
± 2 kV common
mode
< 5% U
T
)
(dip > 95% of U
T
for 0.5 cycles
40% U
T
)
(dip = 60% of U
T
for 5 cycles
70% U
T
)
(dip = 30% of U
T
for 25 cycles
< 5% U
T
)
(dip = 95% of U
T
for 5 cycles
3 A
Electromagnetic environment — directives
G8 Evolution lights use RF energy only for their internal functions.
Their RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
G8 Evolution lights are suitable for use in all establishments other
than domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Electromagnetic environment — guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
T
)
commercial or hospital environment. If the G8 Evolution
T
requires continuous operation during mains interruptions,
it is recommended that it be powered from an
uninterruptible power supply.
T
)
T
T
)
T
T
)
T
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
0138101