CAUTION! Disconnect the unit from
the power supply before cleaning.
• Do not use solvents, chemicals, alcohol, ammonia
or abrasives to clean the housing or control panel.
OPERATING ISSUE
Unit will not turn on.
Cannot zero the
scale or will not zero
when turned on.
Unable to calibrate.
Cannot display
weight in desired
weighing unit.
Error 8.1
Error 8.2
Error 8.3
Error 8.4
Error 8.8
PAGE 10 OF 35
MAINTENANCE
TROUBLESHOOTING
CAUSES
Power cord not plugged in or properly
connected.
Power outlet not supplying electricity.
Battery power used up.
Other failure.
Load on scale exceeds allowable limits.
Load on scale is not stable.
Load cell damage.
LFT security switch set to ON.
Incorrect value for calibration mass.
Unit not set to ON.
Weight reading exceeds Power On/Zero
limit.
Weight reading below Power On/Zero limit.
Weight reading exceeds Overload limit.
Weight reading below Underload limit.
Factory calibration data in EasyConnect
module on load cell cable is not valid
under LFT off status.
CLEANING
• Housing may be cleaned with a cloth dampened
with a mild detergent if necessary.
Check power cord connections. Ensure
power cord is properly plugged into the
power outlet.
Check power source.
Reconnect AC power to charge the
battery.
Service required.
Remove load on scale.
Wait for load to become stable.
Service required.
Set LFT security switch to OFF.
Use correct calibration mass.
Enable unit in the Units Menu.
Remove load from scale. Recalibrate
scale.
Remove red shipping tags from under pan.
Add load to scale. Recalibrate scale.
Reduce load on scale.
Add load to scale. Recalibrate scale.
Calibrate scale.