OPERATING ISSUE
Error 8.9
Err 9.5
CAL E
REF Err
If the troubleshooting section does not resolve your problem, contact Uline Customer Service at 1-800-295-5510.
NOTE: Before using the indicator for the first time, the internal rechargeable battery should be fully
charged for 12 hours. The indicator can be operated during the charging process. The battery is
protected against overcharging and the indicator can remain connected to the AC power line.
PAGE 11 OF 35
TROUBLESHOOTING CONTINUED
CAUSES
Failure to read serial number from
EasyConnect module.
Serial number does not match LFT ON status.
Calibration data not present.
Calibration error. Calibration value outside
allowable limits.
The average piece weight on pan is too
small and less than 0.1d.
1-800-295-5510
uline.com
RECOMMENDATIONS
Break seal or replace original base/
indicator. Service required.
Calibrate scale.
Repeat calibration using correct
calibration weights.
Replace with heavier sample.
0823 IH-10772