5.10 Installing the Control Unit
The control unit can be installed in front of or next to the device.
Procedure
Remove the operating display from the device.
t
Place the control unit in the preferred position (for the dimensions for
t
positioning the control unit, see Chapter "15.1 Dimensions and Weights",
page 79).
5.11
Acclimatization
When a cold device is brought into a warm environment: The temperature
difference can lead to condensation from humidity in the device (moisture
formation). Moisture in the device can lead to malfunctions.
Procedure
Allow the device to acclimatize at the installation site (for the
t
acclimatization duration see Chapter "15.6 Wait Times", page 82).
The device must be disconnected from the power supply during that
time.