Technical Data
15.15 Metrological Data
15.15.1 High-capacity Micro Balance
Models MCA116S-3 | MCA36S-3 | MCA36P-3
Scale interval (d)
Maximum capacity (Max)
Repeatability at 5% load
Standard deviation of the load
values, tolerance
Repeatability at approx. maximum
capacity
Standard deviation of the load
values, tolerance
Standard deviation of the load
values, typical value
Linearity deviation
Tolerance
Typical value
Deviation when load is off-center,
positions according to OImL R76
Test weight
Tolerance
Typical value
Sensitivity drift of +10°C to +30°C
Tare-maximum capacity: Less than
100% of maximum capacity
Accuracy class according to
Directive 2014/31/EU
Verification scale interval (e)
according to Directive 2014/31/EU
Minimum load (Min) according
to Directive 2014/31/EU
Minimum initial weight according to
USP (United States Pharmacopeia),
Chap. 41
Optimum minimum initial weight
Typical stabilization time
Typical measurement time
86
Operating Instructions Cubis
MCA116S-3
Unit
Value
mg
0.002
g
111
mg
0.004
mg
0.01
mg
0.005
mg
0.03
mg
0.02
g
50
mg
0.03
mg
0.02
ppm/K
1
I
mg
1
mg
0.2
mg
1.64
s
3.5
s
8
| MCA Models
®
MCA36S-3
MCA36P-3
Value
Value
0.001
0.001 | 0.01
32
10.1 | 32
0.0015
0.002
0.0025
0.007
0.0018
0.005
0.012
0.015
0.005
0.006
10
10
0.015
0.02
0.006
0.008
1
1
I
I
1
1
0.1
0.1
0.82
0.82
3.5
3.5 | 2.5
10
10 | 6