Descargar Imprimir esta página

J.P. Bowlin Sartorius Cubis II Manual página 801

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 1284
15.8 Recommended Calibration Weight
15.8.1 High-capacity Micro Balance
Models MCA116S-3 | MCA36S-3 | MCA36P-3
Resolution
Supported load
Recommended accuracy class
External test weight
Models MCA66S-3 | MCA66P-3
Resolution
Supported load
Recommended accuracy class
External test weight
15.9 Conditions for isoCAL Function
Possible Conditions for Triggering the isoCAL Function
In the event of a temperature change
After a time interval
After successful leveling
Only conformity-assessed devices: After interrupting the power
supply
Required Conditions for Executing the isoCAL Function
Operating display is in weighing mode (not in menu)
Alphanumeric inputs are not active.
Min. time period without entry on the device
Min. time period with unchanged loading of the weighing pan
Max. loading of weighing pan, high-capacity
MCA116S-3
Unit
Value
mg
0.002
g
111
E2
g
100
MCA66S-3
Unit
Value
mg
0.001
g
61
E2
g
50
MCA36S-3
Value
0.001
32
E2
20
MCA66P-3
Value
0.01 | 0.001
61 | 12
E2
50
Unit
Value
K
1.5
h
12
min
2
min
2
%
2
Operating Instructions Cubis
Technical Data
MCA36P-3
Value
0.01 | 0.001
32 | 10
E2
20
| MCA Models
83
®

Publicidad

loading

Este manual también es adecuado para:

817.332.8116