Page E-18/E-24
AirFlow S2
The AirFlow handpiece hose (18) has a small opening on the bottom of the connector.
If the handpiece hose (18) is not inserted into the coupling (26) properly, air/powder
can escape into the outer covering of the handpiece hose (18). The air/powder mixture
can escape through the small opening. Ensure that this opening does not become
blocked as this could cause the handpiece hose to burst. Should air/powder escape
from the tiny opening, switch the AirFlow S2 off immediately. Then check if the O-rings
(39) of the AirFlow handpiece hose are in good condition and fit properly. Change the
O-rings (39) if necessary. Insert the AirFlow handpiece hose (18) into the coupling (26)
again. If air/powder continues to escape through the opening, the AirFlow handpiece
hose (18) must be replaced.
EMS-Service
Should your AirFlow S2 need additional servicing or repairs, please send it to your
dealer or our EMS service staff. The telephone numbers and addresses of the EMS
service staff covering your area are on the last page of these operating instructions.
The EMS service staff are committed to servicing and repairing your equipment as
quickly as possible.
Only have your AirFlow S2 repaired by EMS or someone authorized by EMS. Only use
original EMS spare parts and accessories. Should the unit be opened without
authorization or damaged due to non-adherence to the operating instructions, EMS
accepts no liability whatsoever. This will also void the guarantee.
Please note - It is best to ship your AirFlow S2 in the original packaging. It protects
your AirFlow S2 against damage during shipping. Before dispatching your AirFlow S2,
please clean and disinfect it and sterilize the Piezon handpiece, instruments, AirFlow
nozzles and AirFlow handpiece barrel as described in the instructions. Always ship
your AirFlow S2 with the complete accessories. When sending your AirFlow S2 directly
to the EMS service department, please include the name and address of your dental
dealer. This simplifies processing for us.