6.5
Measuring weight and height (devices with ultrasonic measuring rod)
1. Press the
field.
Height
2. Enter the height.
3. Press the
key to confirm your entry.
4. Press the
key to clear your entry.
Device mode
Basic
Advanced
Expert
Service
CAUTION!
Injury from falling
Persons with limited mobility may fall when stepping onto the weighing
platform.
► Support people with limited mobility when they step onto the scale.
NOTICE!
Faulty measurement due to reflections
If there are objects or people in the immediate vicinity of the device, in-
correct measurements will result.
► Ensure that there are no objects or people within 0.5 meters of the
front or side of the scale during the measurement procedure.
► Ensure that the device is at least 0.2 meters away from the wall.
► Ensure that the patient is not wearing any kind of hair accessory on
top of their head.
The measurement described in the following is based on the voice guidance
preset at the factory. Information about configuration options can be found here:
è Configuring voice guidance (devices with ultrasonic measuring
rod), page 180.
English
Function available
Operation • 147