Fault/error code
084-297XX-XXX
099-297XX-XXX
10.4
Digital height measurement
Fault/error code
Device in mobile use: Foot sil-
houettes on the weighing
platform not illuminated
Measured value implausible
Measuring heights < 1.22 m:
No measured value shown in
the display
Head slide is hard to move
004-320XX-XXX
192 • seca 655/654; seca 455; seca 257/256/254; seca 453
Cause
Interference caused by reflection
Interference caused by other ultrasonic
emitters
Ambient temperature too high or too
low
Temperature sensor is defective
Voice guidance: Language that does
not support the announcement of mea-
sured results in imperial units is active,
announcement of measured results de-
activated automatically
Cause
Function not available when operated
with
seca 453
battery pack (optional)
Head slide slipped due to greasy sub-
stances on the measuring rod
Digital scanning not available on the
lower telescopic element
Swarf on the column or on the sliding
surfaces of the head slide
No connection between electronic
module and external sensors
Remedy
• Ask people in the vicinity to stay further
away from the device
Increase the distance from other ultrasonic
emitters
Observe ambient conditions for operation,
è General technical
transport, and storage
data, page 205
Inform seca Service
è Switching
• Set metric units
units, page 172
• Activate announcement of measured re-
è Activating/deactivating an-
sults
nouncement of measured results (Re-
sults), page 183
• Select a language that does support the
announcement of measured results in im-
perial units: EN-US, EN-UK, ES-MX, ES-
è Selecting language, page 180
SP
• Activate announcement of measured re-
è Activating/deactivating an-
sults
nouncement of measured results (Re-
sults), page 183
Remedy
If function required, establish power supply
using plug-in power supply unit
• Remove greasy substances
• Do not use hand creams or moisturizing
disinfectants immediately before a mea-
surement
Read off the measured value on the head
slide and enter in patient file manually
è Measuring heights < 1.22 m, page 152
è Hygiene treat-
Clean column
ment, page 187
Inform seca Service