Fault
Required function not avail-
able
Multifunctional display does
not react when keys are
pressed
Multifunctional display re-
mains dark after pressing
10.2
Measuring weight
Fault/error code
Displayed weight is implausi-
ble
Autohold
function cannot be
deactivated
001-272XX-XXX
to
008-272XX-XXX
010-272XX-XXX
Cause
Device mode in which the function is
not provided is active
Device is in an undefined state follow-
ing implausible input
With combinations of devices involving
a handrail and/or measuring rod: Addi-
tional display connected
No connection to weighing platform
No power supply:
• No plug-in power supply unit con-
nected
• On devices with optional battery
pack: Battery pack discharged
Device is in an undefined state follow-
ing implausible input
Multifunctional display defective
Cause
Weighing electronics using outdated
zero point
Weighing electronics defective
Device is connected to a network (in-
tended behavior):
Autohold
activated automatically
Load cell or weight calculation module
defective
Scale has been switched on with too
high a load
Clarify with the administrator/hospital tech-
nician whether the device can be operated
in a different device mode
è Changing device mode, page 160
•
è Functions/device mode, page 203
•
è Restarting the device, page 196
•
• If the error recurs, inform seca Service
è Restarting the device, page 196
•
• If the error recurs, inform seca Service
• Check whether the display cable is con-
nected correctly
• If the error recurs, inform seca Service
• Check whether the plug-in power supply
unit is connected correctly
• On devices with optional battery pack:
è Charging optional battery
pack, page 142
• If the error recurs, inform seca Service
è Restarting the device, page 196
•
• If the error recurs, inform seca Service
Inform seca Service
• Remove the weight from the weighing
platform
• Press
Weight
è Restarting the device, page 196
•
• Wait until main screen is displayed again
Inform seca Service
If necessary, disconnect device from net-
function is
work
Inform seca Service
è Restarting the device, page 196
•
• If the error recurs, inform seca Service
English
Remedy
Remedy
display field
Troubleshooting • 189