Descargar Imprimir esta página

Seca 655 Instrucciones De Uso página 195

Ocultar thumbs Ver también para 655:

Publicidad

Idiomas disponibles

Idiomas disponibles

Fault/error code
010-288XX-XXX
019-288XX-XXX
020-288XX-XXX
022-288XX-XXX
050-280XX-XXX
052-280XX-XXX
053-280XX-XXX
Cause
Update failed
Overcurrent at USB interface of weigh-
ing platform
No connection to
seca connect 103
software
Overcurrent at USB interface of multi-
functional display
Internal device communication error
Communication error between weigh-
ing platform and multifunctional display
Remedy
• Restart update
• If the error recurs, inform seca Service
• Check connected USB device
• If the error recurs, do not use the USB
device
• Repeat measurement
• Check Workflow settings for the
seca connect 103
or
seca analytics 125
software
è Restarting the device, page 196
• If the error recurs, inform seca Service
• Check connected USB device
• If the error recurs, do not use the USB
device
è Restarting the device, page 196
• If the error recurs, inform seca Service
è Restarting the device, page 196
• If the error recurs, inform seca Service
Troubleshooting • 195
English

Publicidad

loading

Este manual también es adecuado para:

654455257256254