11.2
Non-verified scales
11.3
Height measuring devices
200 • seca 655/654; seca 455; seca 257/256/254; seca 453
5. Read off the verification counter.
ð The value must match the number indicated on the verification counter
seal (figure shows example values).
Both numbers have to match for the verification to be valid. If the verification
sticker and the verification counter do not match, the scale must be verified.
Please contact your service partner or seca Service. Once the scale has been
verified, a new, updated verification counter sticker is used to identify the verifi-
cation counter reading. The person authorized to perform verification secures
this verification with an additional seal. The verification counter sticker can be or-
dered from seca Service.
The product needs to be set up carefully and serviced regularly. Depending on
how frequently the product is used, seca recommends servicing at intervals of 3
to 5 years.
CAUTION!
Faulty measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service
or an authorized service partner.
► You can find a service partner in your vicinity at www.seca.com or
by emailing service@seca.com.
The product needs to be set up carefully and serviced regularly. Depending on
how frequently the product is used, seca recommends servicing at intervals of 3
to 5 years.
CAUTION!
Faulty measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service
or an authorized service partner.
► You can find a service partner in your vicinity at www.seca.com or
by emailing service@seca.com.
3
checksum