Descargar Imprimir esta página

Seca 655 Instrucciones De Uso página 152

Ocultar thumbs Ver también para 655:

Publicidad

Idiomas disponibles

Idiomas disponibles

Measuring heights < 1.22 m
Foot silhouettes
152 • seca 655/654; seca 455; seca 257/256/254; seca 453
There is no load on the weighing platform
The upper telescopic element is pushed all the way in
The measuring flap is folded down
1. Release the head slide by pressing on the latch and move the head slide
down onto the lower telescopic element.
2. Fold up the measuring flap until it engages in a horizontal position.
3. Ask the patient to stand under the measuring flap and to adopt the position
below:
• Face the column of the device
• Feet on the illuminated foot silhouettes
• Upright posture: Back and head straight
4. Push the head slide down until the measuring flap is in contact with the pa-
tient's head.
5. Read off the measured result under the head slide as shown in the illustra-
tion.
6. You have the following options for continuing:
► Stand-alone devices: Enter measured result in the patient file manually
► Devices with a network connection: Enter the measured result in the mul-
tifunctional display
the measurement
(è Completing the
7. Ask the patient to step off the weighing platform.
8. Fold down the measuring flap.
9. Push up the head slide until it engages in the latch.
(è Entering height
manually, page 146) and complete
measurement, page 158)