Example(s)
See Also
Fuente de analizador de CD/PMD
Instrucciones de control remoto
Instrucciones IEEE 488.2: Descripción
The table below shows the contents of this
register.
Bit Weight
PON 128
Power ON Enable
N.U. 64
Not used
CMD 32
CoMmanD Error Enable
EXE 16
Execution Error Enable
DDE 8
Device Dependent Error Enable
QRY 4
QueRry Error Enable
N.U. 2
Not used
OPC 1
Operation Complete Enable
A value of 1 in the Enable Register enables the
corresponding bit in the Status Register, a value
of 0 disables the bit. The value of the
<RegisterValue> shall be in the range of 0
through 255.
*ESE 25
where 25 = (bit EXE, bit DDE and bit OPC)
*ESE 0
clears the content of the Standard Event Status
Enable register
*ESE?
*ESR?
*ESE