Instrucciones de control remoto
Instrucciones IEEE 488.2: Descripción
RegisterValue:
The response data syntax for <RegisterValue> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
The <RegisterValue> ranges from 0 through 255.
The <RegisterValue> value expressed in base 2
(binary) represents the bit values of the Standard
Event Status Enable register. See below.
Bit Weight
PON 128
Power ON Enable
N.U. 64
Not used
CMD 32
CoMmanD Error Enable
EXE 16
Execution Error Enable
DDE 8
Device Dependent Error Enable
QRY 4
QueRry Error Enable
N.U. 2
Not used
OPC 1
Operation Complete Enable
*ESE? returns 133
where 133 = (bit PON, bit QYE and bit OPC)
*ESE
*ESR?
*ESE?
Meaning