Package contents
6
1
Refractometer
2
Cable gland and conduit hub
3
Cable (ordered separately)
4
Mounting kit with either standard or high‑pressure clamp (optional)
5
Flow cell (optional)
6
Cooling cover (optional)
Needed tools
• TX20 Torx key
• 9/16 in wrench (for high pressure clamps)
• 22‑mm wrench
• 24‑mm wrench
• Slotted screwdrivers
Choosing the installation location
Follow these instructions for choosing the correct installation location for the process
refractometer. The correct location ensures that the prism stays self-cleaning and that the
lifecycle of the device is not decreased.
Plan enough space around the refractomer for installation, maintenance, and
removal.
• If the temperature varies along the process pipe, select the position with the highest
process temperature. This minimizes the risk of prism coating, because higher
temperature means higher solubility and lower viscosity.
10
1
2
4
5
3
M212866EN-B

Publicidad