Permanently store additional weight (Pre-Tare)
108 • seca 651/650
Device mode
Basic
Advanced
Expert
Service
The
Pre-tare
function can be used to save an additional weight permanently and
subtract it from a measured result automatically, e.g. a flat-rate figure for shoes
and clothing.
Activate
Pre-tare
function
1. Remove the weight from the weighing platform.
2. Press the
key.
PT
3. Enter the value.
4. Confirm the value with the
ð The set additional weight (here: 1.5 kg) is displayed with a minus sign in
front.
ð The messages
NET
5. Ask the patient to step onto the scale.
ð The patient's weight is displayed.
ð The saved additional weight has been deducted automatically.
Deactivate
function
Pre-tare
1. Remove the weight from the weighing platform.
2. Press the
PT
key.
3. Clear the value with the
ð The set additional weight is no longer displayed.
ð The function is deactivated.
Function available
ke.y
and
are displayed.
PT
key.
–
•
•
•