Descargar Imprimir esta página

Seca 651 Instrucciones De Uso página 141

Ocultar thumbs Ver también para 651:

Publicidad

Idiomas disponibles

Idiomas disponibles

11.2
Non-verified scales
5. Read off the verification counter.
ð The value must match the number indicated on the verification counter
seal (figure shows example values).
Both numbers have to match for the verification to be valid. If the verification
sticker and the verification counter do not match, the scale must be verified.
Please contact your service partner or seca Service. Once the scale has been
verified, a new, updated verification counter sticker is used to identify the verifi-
cation counter reading. The person authorized to perform verification secures
this verification with an additional seal. The verification counter sticker can be or-
dered from seca Service.
The product needs to be set up carefully and serviced regularly. Depending on
how frequently the product is used, seca recommends servicing at intervals of 3
to 5 years.
CAUTION!
Faulty measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service
or an authorized service partner.
► You can find a service partner in your vicinity at www.seca.com or
by emailing service@seca.com.
English
3
checksum
Servicing • 141

Publicidad

loading

Este manual también es adecuado para:

650651 7021 099650 1321 009