10.2
Measuring weight
Fault/error code
Displayed weight is implausi-
ble
function cannot be
Autohold
deactivated
to
001-272XX-XXX
008-272XX-XXX
010-272XX-XXX
013-272XX-XXX
016-272XX-XXX
019-272XX-XXX
to
020-272XX-XXX
023-272XX-XXX
024-272XX-XXX
10.3
Data transmission
Fault/error code
The "Date of birth" dialog
window appears after the
key is pressed
134 • seca 651/650
Cause
Weighing electronics using outdated
zero point
Weighing electronics defective
Device is connected to a network (in-
tended behavior):
Autohold
activated automatically
Load cell or weight calculation module
defective
Scale has been switched on with too
high a load
Scale was caused to oscillate and was
unable to determine the zero point
Maximum capacity exceeded
Ambient temperature too high or too
low
One corner of the scale has been
loaded excessively
No GAL value
Cause
Patient ID not scanned
ID scanning not set up
• Remove the weight from the weighing
platform
• Press
Weight
è Restarting the device, page 137
• Wait until main screen is displayed again
Inform seca Service
If necessary, disconnect device from net-
function is
work
Inform seca Service
è Restarting the device, page 137
• If the error recurs, inform seca Service
è Restarting the device, page 137
• If the error recurs, inform seca Service
Ask patient to step off the scale
• Observe ambient conditions for opera-
tion, transport, and storage
technical data, page 145
• Distribute weight evenly
è Restarting the device, page 137
• If the error recurs, inform seca Service
• Inform seca Service
Inform seca Service
• Press the
• Scan patient ID
Check Workflow settings for the
seca connect 103
software
Remedy
display field
è General
Remedy
key
or
seca analytics 125