Fault/error code
After the
key is pressed,
there is no request for patient
or user data
icon appears
icon appears
icon appears
key appears grayed out
key appears grayed out
Autoclear function: An entry
of "0 seconds" is not ac-
cepted; the factory setting
(300 seconds) is suggested
instead
Cause
Patient does not yet have a file in the
EMR System
Device is connected to the
software
seca analytics 125
ID(s) scanned at the start of the mea-
surement
ID scanning not set up
Anonymous
function activated
An obligatory measurement has not
been performed
Scanned ID is invalid
Optional measured value (e.g. height)
has not been determined
Optional ID (e.g. user ID) has been
scanned
There are no data available to be con-
firmed
There are no data available to be
cleared
Implausible input; switching off the
function is not intended in the factory
Remedy
• Enter date of birth and press
again
• Create patient file in the EMR System and
assign measurement
• Enter date of birth
• Press the
key
• Not a malfunction, measured results are
assigned to the patient and saved in the
EMR System
• Check assignment in EMR System
Check Workflow settings for the
or
seca connect 103
seca analytics 125
software
• Not a malfunction, measured results are
sent to the EMR System
• In the EMR System, ensure that mea-
sured results are correctly assigned
• Press the
key
• Perform the missing measurement
Scan valid ID
• Press
key and determine optional
measured value
• Press
key and end measurement
• If desired: Check Workflow settings for
the
seca connect 103
software
seca analytics 125
If desired: Check Workflow settings for the
seca connect 103
or
seca analytics 125
software
• Perform a measurement
• Scan IDs (user/patient)
• Perform a measurement
• Scan IDs (user/patient)
• Accept factory setting
• Manually enter value between 1 and 3600
seconds
Troubleshooting • 135
English
key
or