Item
Device component
WPS button
7
Power supply con-
8
nection
9
Display socket
Spirit level
10
USB interface, multi-
11
functional display
12
Display interface
5.3
Setting up device
96 • seca 651/650
Establishing WiFi connection via WPS
Used to connect the plug-in power supply unit
Used to connect the multifunctional display
Indicates whether the device is horizontal
No function in this device variant, use USB interface on weighing platform
• Used to supply the multifunctional display with power
• Used for data transmission between the multifunctional display and the device elec-
tronics
To achieve accurate measured results, the floor at the setup location must be
level and stable. Soft floors (wooden boards, for example) give under the pa-
tient's weight and falsify the measured result.
1. Place the device on a firm, level surface.
2. Align the device by turning the foot screws.
ð The device is positioned horizontally when the air bubble of the spirit level
is in the precise center of the circle.
CAUTION!
Injury from a lack of stability
If the foot screws are screwed out too far, they may come loose from
the device. The device will then be unstable.
► Screw the foot screws out a maximum of 10 mm.
► If the device cannot be aligned horizontally with the screws
screwed out as far as possible, the setup location is unsuitable. Set
the device up in a suitable location.
3. Tighten the knurled wheels in the direction of the arrow.
ð The foot screws are secured against being adjusted.
Function
Spirit
Foot screw
level
Knurled screw
Air bubble