6. Commissioning, operation
6. Commissioning, operation
6.1 Mechanical connection
In accordance with the general technical regulations for pressure gauges (e.g.
EN
EN 837-2 "Selection and installation recommendations for pressure gauges").
Mount pressure connections in accordance with affixed symbols: ⊕ high pres-
sure, ⊖ low pressure.
Process connections 2 x G ¼ female thread, lower mount, in-line, centre
distance 26 mm, operating position NP 90 (nominal position) per DIN 16257
(i.e. vertical dial), design the threads of the connection shanks preferably in
accordance with EN 837-3 (section 7.3.2).
Prior to the installation of the instrument, clean the measuring lines thoroughly
by tapping and blowing or flushing.
Protect the measuring instruments from contamination and high temperature
changes!
Mount the instrument free from vibration and make alignment so that it is easy
to read. It is recommended that a shut-off device is interposed between the
pressure tapping point and the instrument, which will enable the replacement
of the instrument and a zero point check while the plant is running. Protect the
instrument against coarse dirt and wide fluctuations in ambient temperature.
For sealing the connections, use flat gaskets, lens-type sealing rings or WIKA
profile seals. Connections with LH-RH adjusting nut or union nut enable the po-
sitioning of the instrument for better readability. For the mounting or dismount-
ing of the instrument, the required force must not be introduced via the case,
but only on the spanner flats of the connection nozzle!
Wall mounting
Mounting using three integrally cast mounting lugs